Robert Bidigare
NBP97-3
Methodology notes

A comparison of the TURNER-determined chlorophyll a concentrations with the HPLC-determined TCHLA concentrations (monovinyl chlorophyll a + monovinyl chlorophyllide a; units = ng Chl a equivalents/L) was performed for Ross Process Cruise III (NBP9703). A good agreement between these two methods was observed (see below).

Results of geometric mean regression analysis (reduced major axis):

Y = HPLC TCHLA (monovinyl chlorophyll a + monovinyl chlorophyllide a), units = ng Chl a equivalents/L

X = TURNER chlorophyll a (it is necessary to convert the Turner Chl a concentrations in the AESOPS data base from mg/m3 to ng/L by multiplying concentrations by 1000)

HPLC TCHLA = TURNER*(0.7601) + 5.9113 (r = 0.8489, n = 72)